Delayed vasopressor delivery

Educational use only. Not clinically reviewed

58-year-old woman for Vasopressor infusion-path and startup review. About 8 simulated minutes.

What you will practise

  • Recognize that a running pump command does not prove vasopressor delivery to the patient.
  • Trace the declared syringe, pump, tubing, connections, downstream volume, carrier, catheter, and patient.
  • Classify the fixed pattern as delayed delivery from dead-space and startup mechanics while keeping other causes open.
  • Activate nursing, pharmacy, and local safe-start or changeover protocol without flushing concentrated drug into the patient.
  • Reassess delivery evidence, pressure, perfusion, and the ongoing shock plan.

Review and sources

Not clinically reviewed. No clinician has signed this scenario.

  • Weiss M, van der Eijk A, Lönnqvist PA, Lucchini A, Timmerman A. 10 clinical tips for advancing patient safety when using syringe pump systems for microinfusion intravenous drug therapy. Eur J Anaesthesiol. 2023;40:387-390. PMID:37132300.
  • Lovich MA, et al. Infusion system carrier flow perturbations and dead-volume: large effects on drug delivery in vitro and hemodynamic responses in a swine model. Anesth Analg. 2015;120:1255-1263. PMID:25811259.
  • Baeckert M, Batliner M, Grass B, et al. Performance of modern syringe infusion pump assemblies at low infusion rates in the perioperative setting. Br J Anaesth. 2020;124:173-182.
  • Neves EB, et al. Start-up delay in syringe infusion pumps with different rates and priming techniques of intravenous sets. Rev Gaucha Enferm. 2022;43:e20210071.